Thursday, 24 July 2014

REDUCING TURN AROUND TIME IN CLINICAL LABORATORY



REDUCING TURN AROUND TIME IN CLINICAL LABORATORY
Diagnostic responsiveness is critical to providing optimal service in high volume patient care; this is particularly important in emergency and operating rooms. With 60-70% of the required information on a patient’s chart coming from laboratory test results, the demand for quick service is translated into aggressive time requirements for cycle times of ordered tests (Holland, Smith, & Blick, 2005). (1) In clinical laboratories, the time from when a test is ordered to when the results are verified is defined as the turnaround time (TAT). The effects of TAT have been studied to a high extent, with correlations being drawn between emergency department treatment and length of stay (Hawkins, 2007). (2) As a result, TAT is often considered the most significant measure of a laboratory’s service and is used by many clinicians to judge its quality. Along with accuracy and reliability, timely reporting of laboratory test results is now considered an important aspect of the services provided by the clinical laboratory. Whether or not, faster turnaround time can make any medical difference, patients and their physicians want reports as rapidly as possible. There is increasing pressure from clinicians to report results rapidly. Even though there are only sparse data, timeliness in reporting of laboratory results undoubtedly affects clinician and patient satisfaction as well as length of hospital stay. Improving turnaround time (TAT) is a complex task involving education, equipment acquisition, and planning. All the steps from test ordering to results reporting should be monitored and steps taken to improve the processes. Poor core laboratory performance that causes delays in diagnosis and treatment is an impediment to optimal patient care, particularly in high-volume patient care areas such as the emergency department (ED). Most easy way to reduce Turn-Around-Time is to purchase the high throughput analyzer. However, if the Sample Transportation System has enough intelligence and performance, it is possible to minimize the investment for analyzers. In other words, it will be an over investment if the user needs larger scale of analyzer to cover the poor performance of transportation system.

Definitions
The Clinical Laboratory Standards Institute (CLSI, Wayne, PA) is working on a document (GP35) aimed at the "development and use of quality indicators for process improvement and monitoring of laboratory quality." Review of the literature shows a number of approaches in defining TAT. In general, it has been categorized by test (individual assays or panels), priority (stat or routine) and patients served (inpatient, outpatient or ED).
Lundberg described the path of workflow from the ED to the laboratory as a series of nine steps: identification, order, collection, ­transportation, preparation, analysis, reporting, result interpretation and action for treatment. Howanitz, who has published widely on CAP survey results, has suggested that TAT be defined from the time physicians place the laboratory orders to the time the test results are available to the clinician. He suggests that the expected turnaround time goals should be expressed as a percentage of all results completed within a given time interval. However, to maintain a successful ED TAT program there is a need to consider both good metrics and the engagement of staff members to resolve outliers as they occur on real time. We define ED TAT as follows:
1. ED Performance (ED Interval)
Monitoring the ED's timeliness of performance during the pre-analytical interval order-to-receipt.
a) START: The time physicians place orders for laboratory examinations, including clerical time needed to receive and log laboratory orders in the hospital information system.
b) MIDDLE: It continues with the time needed to collect and transport patient samples to the laboratory.
c) END: When the laboratory assistants receive and log all collected samples in the LIS.

2. Lab Performance (Lab Interval)
a) Monitoring the laboratory's timeliness of performance during the analytical interval receipt-to-report.
b) START: When samples get delivered to the medical technologists work cells for processing.
c) MIDDLE: It continues with the time it takes to prepare, centrifuge and process patient samples by instrumental analysis. It includes the time needed for dilutions, retesting and notification of critical results.
d) END: When medical technologists verify and broadcast results to the medical staff in the laboratory information system.

3. Total Testing Cycle (ED + Lab Intervals):
Monitoring the collective effort by both the ED and lab to complete the total testing cycle, order-to-report. That is, the time it takes both to finalize all orders.

4. ED Outliers:
Delays found in the ED's preanalytical phase. Controllable and uncontrollable delays during ordering, collection and transportation of laboratory samples by ED personnel are included. A 10 percent allowance is included for orders that may take over 45 minutes to be processed by the ED.

5. Laboratory Outliers:
Delays found in the laboratory's analytical and post-analytical phases. Expected and unexpected delays during specimen receiving, processing, testing and reporting by the laboratory staff are included. A 10 percent allowance is included for samples that may take over 45 minutes to be completed by the laboratory

Outlier analysis is commonly used in industrial quality control as a method of determining reasons for system failure. In the clinical laboratory, outlier rates are expressed as the percent of total ­laboratory examinations failing to meet targeted reporting times. Improving TAT requires education of a wide variety of individuals, long-term planning, and completion of innumerable tasks. Small investments in the clinical laboratory resources may improve TAT and greatly improve clinicians’ efficiency, as well as help reduce required days of hospitalization for patients. Overall cost reduction, however, may be difficult, if not impossible, to prove. With limited exceptions,studies to date fail to show that decreased TAT improves thelength of hospital stay or patient care.(3,4,5,6)  Practitioners needsto become more efficient. From the clinicians’ viewpoint, it is easy to see the benefits of rapid return of results. With the appropriate information available, laboratory results can be explained to the patient and treatment adjusted all in one encounter, thus increasing clinician efficiency and patient satisfaction. If laboratory results provide essential information for patient diagnosis and treatment, it follows that more timely results will improve patient care. Patient outcomes undoubtedly are affected by delays in diagnosis. (7) It is important to remind those who allocate resources that laboratory results must be available not only for diagnostic use but
also before many treatments and procedures can be implemented. Thus, despite the lack of data, it is reasonable to assume that timeliness of laboratory results affects physician efficiency and hospital length of stay. Therefore, monitoring and enhancing timeliness of results reporting are fundamental to laboratory quality improvement.

TAT Benchmarks
Laboratory professionals generally believe intralaboratory TATs of up to 60 minutes are appropriate; clinicians do not agree.(8,9) Clinicians consider TAT from the time the test is ordered to results reporting, whereas laboratory professionals usually use specimen receipt to reporting of results as the TAT.(10) Although it is difficult to monitor each of the steps from ordering a test to reporting the results, and often laboratory professionals are not in control of many of the processes, it is
important to view TAT from the clinicians’ viewpoint. TAT should be monitored on a regular basis, not only for the mean TAT, but also for results reported well beyond the average TAT (ie, outliers). Various measurement parameters have been used to express TAT, including proportion of acceptable results.(11) Mean and median TATs are not affected significantly by outliers, and, thus, they are not good statistical indicators for laboratories with good performance that want to improve further.16 Identifying outliers and looking for the root cause of these problems is an excellent approach to understanding and eventually eliminating untimely reporting of results. Another approach is to monitor inpatient test availability for morning rounds.4 Through the College of American Pathologists Q-Probes and Q-Tracks programs, national databases on a number of TAT parameters have become available. Recently, the College of American Pathologists offered 2 Q-Tracks TAT monitors: Stat Test Turnaround Time Outliers and Morning Rounds Inpatient Test Availability.(12)
At present, TATs for most stat clinical laboratory tests are less than 1 hour and 2 working days for most routine surgical pathology cases.18-20 Using report inquiry for CBC count reports as an indicator, data show that most reports for inpatients and outpatient tests ordered stat are requested within 4 hours.(13,14) Others make the point that timeliness in reporting early morning routine clinical laboratory test results is an important parameter. It is important to choose TAT goals that lead to improved patient care and clinician efficiency and to improved satisfaction for both patients and clinicians. Ideally, all common laboratory tests should be reported as rapidly as possible by methods yielding high quality results, and this currently means an hour or less from order entry to results reporting under optimal conditions. For effective management of resources and improvement in patient satisfaction, it is especially important to report outpatient results promptly.
The resolution of delays, measured in terms of outliers, is one of the most important benchmarks in meeting ED TAT goals. Resolving delays caused in ED TAT reflect efforts to provide diligent laboratory test results to improve patient care. This may be achieved by:
Ø  raising employees' awareness in tracking and resolving delays as they happen in real time,
Ø  motivating employees to achieve daily ED TAT goals by decreasing the number of outliers,
Ø  keeping employees informed with daily ED TAT reports,
Ø  recognizing staff members when ED TAT daily goals are met and
Ø  having an action plan to resolve delays that could be avoided or controlled.

Every hospital laboratory has its own ways of monitoring ED TAT. TAT data collection and ­outcome interpretation varied. This may be due, in part, to:
Ø  the different time frames used in monitoring ED TAT-in real time, daily, weekly or monthly;
Ø  the selection of different lab activity test menus used to monitor ED TAT;
Ø  the types of equipments used to examine patient samples;
Ø  different hospital and laboratory information systems used in tracking ED TAT; and/or
Ø  types of software products used to tabulate TAT data.

Strategy:
1. Define the services provided, such as urine analysis or blood analysis, and measure current turn times overall. Ask the customer to define her expectation and analyze the gap. Use the smallest measurement increment available, such as seconds or minutes. For example, the customer wants a turn time of 30 minutes for a hemoglobin blood test, and right now you are yielding 60 minutes. The objective in this case would be to reduce current turn time by 30 minutes.

2. Remove variation first. Define each major step of the process such as customer intake, computer update, sample drawn, sample labeled, sample tested, testing validated, results recorded, customer notified, and customer billed. Visualize entire process by creating a top-level flow chart. Instruct employees to handle every request with a uniform first-in/first-out process so that when a problem occurs it receives visibility. Focus on flow and do not allow employees to put work aside for clarification as these "exception situations" cause 80 percent of the variation in the outputs. Concentrate efforts on ensuring a consistent process even if the process isn't yet yielding the results you need.

3. Measure each sub-step in the process to see how long it takes. In this case, we find we have 5 minutes for customer intake, 20 minutes waiting, 7 minutes for blood draw, 1 minute for labeling, 2 minutes for analysis, and 20 minutes waiting before clerk notifies customer of results. In this case, the waiting or queue time is what is eating up turn time. Adjust process to minimize wait time. To do this, identify areas that exhibit waste, such as overproduction (making things unnecessarily like copies), delays (waiting), transportation (moving data or equipment), wasted motion (walking, non-value added keystrokes), inventory (mismanaged, not enough, too much) and making defective products (adding work to something already defective). Develop ways to streamline process and communication flow by re-timing or moving work, consolidating tasks, getting rid of non-valued activities, and adding machines or personnel.

4. Benchmark competitors' laboratories to compare how they handle clinical laboratory services. Investigate opportunities to apply new software, new technologies and methods to improve accuracy and responsiveness. Make sure measurement systems are consistent between provider and customer. For example, the customer may measure the process up to the point that they receive the results whereas a provider may measure the process up to the point of payment. Examine the real needs of the desire to drive down turn times and address these specific areas each time.

5. Control the process by measuring inputs to make sure steps within the process remain consistent. Investigate anomalies and figure out why problems occur. Be proactive against variation by improving training and hiring methods. If one technician takes longer than another, find out why. For example, during the blood draw stage, the patient is brought to the room, a tourniquet is wrapped around his arm, the technician takes the needle in one hand and a swab in another. After swabbing the area, the technician penetrates the vein with a needle, and releases the tourniquet and then draws the blood. They take the pre-formatted label and affix it to the blood sample and set it in the test area. Another technician talks incessantly, draws the blood the same way but instead of labeling and putting the blood sample in the test area, they walk the customer out the door, and forget to put the sample in the test area until an hour later. This would be the variation that needs to be eliminated to improve the process.

6. Shift the bell curve slowly and methodically, incrementally. If the goal is to reduce turn times by 50 percent or 30 minutes overall, then work to reduce seconds off each step of the process. Set realistic upper and lower control limits by examining current yields. For example, if the blood draw process takes 7 minutes on average, set the lower limit to 60 seconds and the upper limit to 420 seconds. Any time a blood draw takes longer than the 420 seconds, figure out why and eliminate the problems. Each time a task within the process is perfected to take less time, the overall turn time will be reduced, getting you closer and closer to your goal.

Regular audit of data helps in the evaluation of the efficiency of the laboratory and hence corrective measures taken accordingly would be helpful in providing better service to the physicians and patients
Thus, many have predicted that the use of point-of-care testing will increase, but the question remains whether point of- care testing is the most efficient and effective means of producing test results.(15) Currently, technologic advances are occurring that can make large improvements in central laboratory TAT a reality. Now is the time to improve timeliness not only as a strategy but also as a duty to all of our customers, clinicians and patients alike. There are constant pressures to make trade-offs, to compromise, and to change strategy. Timeliness of results reporting is too important an issue to fall prey to these pressures.

1)     Hawkins, R. C. (2007). Laboratory turnaround time. Clinical Biochemist Reviews, 28(4), 179-194.
2)     Holland, L. L., Smith, L. L., & Blick, K. E. (2005). Reducing laboratory turnaround time outliers can reduce emergency patient length of stay. American Society for Clinical Pathology, 124, 672-674.
3)     Steindel SJ, Jones BA, Howanitz PJ. Timeliness of automated routine laboratory tests: a College of American Pathologists Q-Probes study of 653 institutions. Clin Chim Acta. 1996; 251:25-40.
4)     Steindel SJ. Timeliness of clinical laboratory tests. Arch Pathol Lab Med. 1995;119:918-923.
5)     Saxena S, Wong ET. Does the emergency department need a dedicated stat laboratory? Am J Clin Pathol. 1993;100:606-610.
6)     Kost GJ. Connectivity: the millennium challenge for pointof-care testing. Arch Pathol Lab Med. 2000;124:1108-1110.
7)     Kenagy JW, Berwick DM, Shore MF. Service quality in health care. JAMA. 1999;281:661-665.
8)     Hilborne LH, Oye RK, McArdle JE, et al. Evaluation of stat and routine turnaround times as a component of laboratory quality. Am J Clin Pathol. 1989;91:331-335.
9)     Howanitz PJ, Cembrowski GS, Steindel SJ, et al. Physician goals and laboratory test turnaround times. Arch Pathol Lab Med. 1993;117:22-28.
10)  Weinstein S. Quality in pathology laboratory practice. J Qual Clin Pract. 1995;15:121-126.
11)  Valenstein PN, Emancipator K. Sensitivity, specificity, and reproducibility of four measures of laboratory turnaround time. Am J Clin Pathol. 1989;91:452-457.
12)  College of American Pathologists. Q-Tracks, Q-Probes [brochure]. Northfield, IL: College of American Pathologists; 2001.
13)  Burke MD. Laboratory test turnaround time and the needs of medical care. Am J Clin Pathol. 1997;108:367-368.
14)  22. Winkelman JW, Tanasijevic MJ, Wybenga DR, et al. How fast is fast enough for clinical laboratory turnaround time? measurement of the interval between result entry and inquiries for reports. Am J Clin Pathol. 1997;108:400-405.
15)  Kane B. Point-of-care testing: instant gratification? Ann Intern Med. 1999;130:870-872.

Tuesday, 22 July 2014

STATUS OF CAPILLARY GLUCOSE ESTIMATION IN POINT-OF-CARE TESTING



STATUS OF CAPILLARY GLUCOSE ESTIMATION IN POINT-OF-CARE TESTING
PRINCIPLE:
Point of Care Testing (POCT) refers to those analytical patient testing activities provided within the hospital, but performed outside the laboratory. Activities of the POCT program must comply with all current standards for laboratory accreditation, regardless of the scope of testing. Designated laboratory personnel have centralized the coordination of the POCT program. Glucose meters are widely used in many areas including hospital wards and clinics, general practice, operating theatres, ambulances, pharmacies, aboriginal health care centres and by patients at home for self-monitoring. Such testing involves the use of a glucose measuring meter, a specific blood glucose test strip and appropriate quality control material. This will be referred to here under as the Glucose system. Glucometers are classified into Class II devices according to US FDA which means they should fulfill following criteria
•Require 510(k) clearance prior to marketing
•Substantial equivalence to predicate
•FDA evaluates intended use, performance, labeling

FDA clears glucose meters for the following indications:
─Quantitative measurement of glucose in whole blood (e.g., capillary, venous, arterial)
─Intended for self testing outside the body (in vitro diagnostic use) by people with diabetes
─ As an aid in monitoring the effectiveness of diabetes control program
─ NOT intended for the diagnosis of or screening for diabetes
-     Testing sites - fingertip and alternative sites (e.g. forearm, palm, calf, thigh)
– A few are cleared for use on neonates


Majority of meters are designed and validated for OTC use
v  The majority of meters used in hospitals are OTC devices
v  CLIA waived by regulation – used anywhere in the professional healthcare setting
v  ISO 15197 accuracy criteria for OTC use
v  User is not the intended user

Following are taken into account when using these devices:
 The limitations of the system when used in specific settings such as critical care units, obstetric wards and neonatal units
 The effects of physiological and pathological interferences such as haematocrit, various drugs and various intravenously administered solutions
 Other potential interfering substances
The analytical performance (precision and bias) of glucose systems should be designed to match the clinical application of the test. This document outlines recommendations aimed at improving the use of glucose systems for patient care.

Recommendations from Australian Association of Clinical Biochemist
1. All glucose systems used by health professionals for patient management should participate in an EQA program. Review and communication of the performance of each meter in such programs should become a routine practice for each site.
2. It is highly desirable to run a quality control sample daily. This ensures ongoing validation of the test strips.
3. It is recommended that glucose systems which have features designed for professional use are implemented in acute hospital facilities, not systems designed for home use. This will allow results to be downloaded to patient information systems so results are part of the patient record. This ensures more effective use of results in patient management. Improved patient management for diabetes will justify increased capital outlay for professional glucose meters.
4. Glucose systems used in hospital facilities should allow patient identification as well as the capacity to be connected to other systems and/or networks.
5. The health care professional managing the patient should decide on the glucose system most suitable to the patient setting, taking into account the relevant professional scientific advice available for each device.
6. Glucose estimations performed in Intensive Care Units should be performed on a blood gas machine or device of similar analytical performance.
7. It is highly desirable that the tolerance for analytical bias (accuracy) and imprecision of glucose systems in other hospital areas should be less than 10% variance from the true value.
8. Glucose systems accuracy should be confirmed by an accredited laboratory method at least once every 12 months. If analytical performance is less than that required, the glucose system should be replaced.
9. A standardised training package for health professionals and patients is essential. As a minimum health care professionals should complete an online competency program such as that provided by The Australian Point of Care
Practitioner’s Network (APPN) which is currently funded by the Australian Government through the Quality Use of Pathology Program (www.appn.net.au).
10. It is recommended that all incidents and adverse events are reported to TGA through the IRIS system (http://www.tga.gov.au/safety/problem-device.htm)
11. The performance of capillary blood glucose system should be validated by an appropriately accredited clinical laboratory. Such validation should follow the protocol published by the AACB PoCT Committee. This validation report must be a minimum requirement prior to the introduction of a ward based glucose testing system in a hospital.
12. Glucose meters designed for self-monitoring at home should not be used to monitor tight glycaemic control.
13. It is desirable to assess performance glucose systems designed for self-monitoring in the hands of patients as well as health care professionals.


Confirmation Testing:
Critical (panic) values obtained on glucometer are required to be followed up with confirmation laboratory testing. This is accomplished by nursing personnel either: a. drawing a specimen and keying in a POC Confirmation Glucose or
b. if regular lab is being or has been drawn immediately prior to test, the nursing personnel will notify the lab of the critical POC value to expedite confirmation testing of the glucose. If the lab office is notified, they must pass the information to Routine Chemistry, Trauma Lab, or Outpatient Lab, whichever will be performing the test.

All confirmation testing is to be handled as high priority. Confirmatory test results are called to the appropriate nurse or physician in charge of the patient. This will provide verification of POC values or initiate exchange of the Inform monitor by nursing personnel to ensure quality patient care.
A number of different studies have examined the accuracy of bedside glucose meters in various patient populations with either hypotension, poor peripheral perfusion, or edema. Two studies done on patients in the emergency department and ICU with hypotension (systolic blood pressure less than 80 mm Hg) both found that capillary glucose measurement systematically underestimated venous glucose in this patient population. (1,2) In contrast, another study performed on 75 patients with systemic hypoperfusion (defined as systolic blood pressure less than 90 mm Hg or requirement for vasoactive agents) found relatively good agreement between capillary whole blood and arterial whole blood glucose, with 95% limits of agreement of approximately [+ or -]30 mg/dL. (3) Finally, a recent study comparing capillary whole blood to venous plasma glucose in patients with poor peripheral perfusion related to vasopressor use or peripheral edema found that in both categories of patients (poor perfusion and peripheral edema), capillary whole blood glucose systematically overestimated venous plasma glucose. Even in this study, however, almost all results agreed within approximately 2 mmol/L (36 mg/dL) glucose. (4)
In recent years, numerous studies have been published analyzing the accuracy of glucometers specifically in the setting of neonatal hypoglycemia. Ho, et al. (5) reported on the sensitivity and negative predictive value of 5 glucometers in detecting neonatal hypoglycemia. They found that not even 1 of the 5 was able to meet the ADA standards, whereas 2 of the devices were able to meet the NCCLS standards. Khan, et al. (6) compared 7 glucometers and reported agreement between glucometer readings in the hypoglycemic range but found wide discrepancy in the correlation between reference and POCT devices both in the hypo and hyperglycemic range to the tune of 60%. The study by Ngerncham, et al. (7) appearing in this volume of the journal too has used an elegant split sample design and meticulously compared OneTouch SureStep Hospital Test Strips (photometric glucose oxidase system) with a Nova StatStrip (modified glucose oxidase based amperometric system) using Roche Modular P 800 for the reference laboratory measurement. Another recent study reported good correlation as well as recommended their use in neonatal clinical practice (8). All these studies highlight that POC devices may be used as screening devices for neonatal hypoglycemia, but confirmation of hypoglycemia with laboratory measurement of plasma glucose is still crucial.
It needs to be realized that laboratory estimation too is fraught with preanalytical (sample collection, transport and physiological factors) errors. POCT devices are prone for both pre-analytical and analytical errors (precision of the device being used). The comparison of POCT devices with laboratory sample which has been poorly processed can lead to erroneous estimation of discrepancy where little or none may exist. Any further studies on comparison of POCT devices with reference standard should focus on comparing improved second generation POCT devices with a reference laboratory device taking utmost care to ensure precise laboratory estimation without any processing delays and fall in glucose secondary to glycolysis.

Australian Diabetes Educators Association
The Australian Diabetes Educators Association (ADEA) promotes the reliable and accurate use of blood glucose meters within the health care setting and in diabetes self management.
ADEA recommends:
 Blood glucose meters not be used as a method of screening for diabetes.
 Testing at the point of care may be appropriate in defined circumstances, such as in remote indigenous communities where laboratory testing is unavailable and postponement of treatment would be potentially harmful for the individual. The meter used at the point of care must have a vigorous quality improvement program in use.
 Outside of the acute clinical setting, blood glucose meters only be used to monitor blood glucose levels in people with a confirmed diagnosis of diabetes.
 Appropriateness of self blood glucose monitoring be assessed on an individual basis, taking into consideration the person’s disease and co-morbidity status, age, culture, dexterity and physical and intellectual capabilities, identified glycaemic targets, current medication regimen, potential confounders that may interfere with the accuracy of results obtained, and level of motivation.
 All people with diabetes using insulin therapy are encouraged to perform blood glucose monitoring.
 Individuals using blood glucose meters have access to a health professional deemed competent in the use of the meter.
 Further diabetes education is essential to ensure self management behaviours are underpinned by a sound knowledge of how dietary intake, physical activity, medication, stress and illness all interact to affect blood glucose levels.
 Although some blood glucose meters allow the measurement of glucose levels from small samples of blood from the forearm and other sites, blood taken from the capillary bed of the fingertip is the preferred sample, particularly when blood glucose levels are changing rapidly.
 Health professionals only use blood glucose meters after successfully completing an education program that results in the attainment of competency in meter operation, control testing and problem solving.
 Quality improvement practices are implemented within clinical settings to ensure both blood glucose monitoring equipment and operators meet high standards of performance and process.
 All health services using blood glucose meters provide:
Ø  a well-defined policy and procedure
Ø  a training program for personnel performing the tests
Ø  quality improvement procedures
Ø  regular equipment maintenance
Ø  external auditing of meters
Ø  appropriate lancet devices that meet infection control considerations.

 In certain clinical situations different strip technology may be more appropriate than others, such as in the care of patients using icodextrin dialysate solution, or patients receiving intravenous preparations containing maltose. In these situations it is essential to consult product information and/or the manufacturer of the glucose meter and test strip to ensure icodextrin and maltose do not interfere with blood glucose readings.

 Blood glucose measurement using portable blood glucose meters not be used in isolation when evaluating the glycaemic control of people with diabetes.

Substances or Conditions That May Contribute to Total Analytical Error in Consumer and Hospital Glucose Meters Interferant Reference
ü  Hematocrit
ü  Ascorbate
ü  Maltose
ü  Galactose or Xylose
ü  Acetaminophen
ü  Partial Pressure of Oxygen
ü  Dopamine
ü  Mannitol
ü  Altitude or hypoxia

Numerous studies have been published reporting the performance of blood glucose monitoring systems. The results of these studies are often conflicting. Correct evaluation is, however, complex, and the apparent contradiction of results creates confusion. One recent article (9) provides an overview of frequently made errors in the evaluation of devices and develops an easy-to-use checklist to verify the quality of such studies. This checklist was used with 20 representative studies selected from the literature between 2007 and 2012. The process revealed that limitations on the designs and methods of studies assessing the performance of blood glucose systems are common.
The use of the accuracy checklist showed that 40% of the studies showed clear nonconcordance with ISO 15197. The use of the interference checklist showed that only half of publications were in good agreement with the quality checks. The author concluded that many evaluations are performed poorly, and often present questionable conclusions due to the fact that few publications adhere to international guidelines. Others have also assessed the quality of glucose monitor studies and came to similar conclusions.
In the article by Mahoney and Ellison,(10)52 reports were reviewed, none of which conformed to all 38 STARD and CLSI recommendations. Future studies evaluating glucose monitoring systems should be carefully designed and should follow published recommendations to assess the quality of glucose measuring devices

References:
1. Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S. Fingerstick glucose determination in shock. Ann Intern Med. 1991;114:1020-1024.
2. Sylvain HF, et al. Accuracy of fingerstick glucose values in shock patients. Am J Crit Care Med 1995;4:44-48.
3. Kulkarni A, et al. Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med. 2005;31:142-145.
4. Kanji S, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Am J Crit Care Med. 2005;33:2778-2785.
5. Ho HT, Yeung WK, Young BW. Evaluation of point of care devices in the measurement of low blood glucose in neonatal practice. Arch Dis Child Fetal Neonatal Ed. 2004;89:F356-9.
6. Khan AI, Vasquez, Y, Gray J, Wians FH, Kroll MH. The variability of results between point of care testing glucose meters and the central laboratory analyzer. Arch Pathol Lab Med. 2006; 130: 1527-32.
7. Ngerncham S, Piriyanimit S, Wongsiridej P, Inchgarm L, Kitsommart R, Vutrapongwatana P, et al. Validity of two point of care glucometers in the diagnosis of neonatal hypoglycemia. Indian Pediatr. 2012;49: 621-5.
8. Makaya T, Memmott A, Bustani P. Point-of-care glucose monitoring on the neonatal unit. J Pediatr Child Health. 2012;48:342-6.
9. Thorpe GH. Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems. Diabetes Technol Ther. 2013;15(3):253-259.
10. Mahoney J, Ellison J. Assessing the quality of glucose monitor studies: a critical evaluation of published reports. Clin Chem. 2007;53(6):1122-1128.