STATUS OF
CAPILLARY GLUCOSE ESTIMATION IN POINT-OF-CARE TESTING
PRINCIPLE:
Point of Care Testing (POCT) refers to
those analytical patient testing activities provided within the hospital, but
performed outside the laboratory. Activities of the POCT program must comply
with all current standards for laboratory accreditation, regardless of the
scope of testing. Designated laboratory personnel have centralized the
coordination of the POCT program. Glucose meters are widely used in many areas
including hospital wards and clinics, general practice, operating theatres,
ambulances, pharmacies, aboriginal health care centres and by patients at home
for self-monitoring. Such testing involves the use of a glucose measuring
meter, a specific blood glucose test strip and appropriate quality control
material. This will be referred to here under as the Glucose system.
Glucometers are classified into Class II devices according to US FDA which
means they should fulfill following criteria
•Require 510(k) clearance prior to marketing
•Substantial equivalence to predicate
•FDA evaluates intended use,
performance, labeling
FDA clears glucose meters for the
following indications:
─Quantitative measurement of glucose
in whole blood (e.g., capillary, venous, arterial)
─Intended for self testing outside the
body (in vitro diagnostic use) by people with diabetes
─ As an aid in monitoring the
effectiveness of diabetes control program
─
NOT intended for the diagnosis of or screening for diabetes
- Testing sites - fingertip and alternative
sites (e.g. forearm, palm, calf, thigh)
– A few are cleared for use on
neonates
Majority of meters are designed and
validated for OTC use
v
The majority of meters used in
hospitals are OTC devices
v
CLIA waived by regulation – used
anywhere in the professional healthcare setting
v
ISO 15197 accuracy criteria for OTC
use
v
User is not the intended user
Following are taken into account when
using these devices:
The limitations of the system when
used in specific settings such as critical care units, obstetric wards and
neonatal units
The effects of physiological and
pathological interferences such as haematocrit, various drugs and various
intravenously administered solutions
Other potential interfering
substances
The analytical performance (precision
and bias) of glucose systems should be designed to match the clinical
application of the test. This document outlines recommendations aimed at
improving the use of glucose systems for patient care.
Recommendations from Australian Association of Clinical Biochemist
1. All glucose systems used by
health professionals for patient management should participate in an EQA
program. Review and communication of the performance of each meter in such
programs should become a routine practice for each site.
2. It is highly desirable to run a
quality control sample daily. This ensures ongoing validation of the test
strips.
3. It is recommended that glucose
systems which have features designed for professional use are implemented in
acute hospital facilities, not systems designed for home use. This will allow
results to be downloaded to patient information systems so results are part of
the patient record. This ensures more effective use of results in patient
management. Improved patient management for diabetes will justify increased
capital outlay for professional glucose meters.
4. Glucose systems used in hospital
facilities should allow patient identification as well as the capacity to be
connected to other systems and/or networks.
5. The health care professional
managing the patient should decide on the glucose system most suitable to the
patient setting, taking into account the relevant professional scientific
advice available for each device.
6. Glucose estimations performed in
Intensive Care Units should be performed on a blood gas machine or device of
similar analytical performance.
7. It is highly desirable that the
tolerance for analytical bias (accuracy) and imprecision of glucose systems in
other hospital areas should be less than 10% variance from the true value.
8. Glucose systems accuracy should
be confirmed by an accredited laboratory method at least once every 12 months.
If analytical performance is less than that required, the glucose system should
be replaced.
9. A standardised training package for health professionals
and patients is essential. As a minimum health care professionals should
complete an online competency program such as that provided by The Australian
Point of Care
Practitioner’s Network (APPN) which
is currently funded by the Australian Government through the Quality Use of
Pathology Program (www.appn.net.au).
10. It is recommended that all
incidents and adverse events are reported to TGA through the IRIS system
(http://www.tga.gov.au/safety/problem-device.htm)
11. The performance of capillary
blood glucose system should be validated by an appropriately accredited
clinical laboratory. Such validation should follow the protocol published by
the AACB PoCT Committee. This validation report must be a minimum requirement
prior to the introduction of a ward based glucose testing system in a hospital.
12. Glucose meters designed for
self-monitoring at home should not be used to monitor tight glycaemic control.
13. It is desirable to assess performance glucose systems
designed for self-monitoring in the hands of patients as well as health care
professionals.
Confirmation Testing:
Critical (panic) values obtained on glucometer
are required to be followed up with confirmation laboratory testing. This is
accomplished by nursing personnel either: a. drawing a specimen and keying in a
POC Confirmation Glucose or
b. if regular lab is being or has been drawn
immediately prior to test, the nursing personnel will notify the lab of the
critical POC value to expedite confirmation testing of the glucose. If the lab
office is notified, they must pass
the information to Routine Chemistry, Trauma Lab, or Outpatient Lab, whichever
will be performing the test.
All confirmation testing is to be handled as high priority.
Confirmatory test results are called to the appropriate nurse or physician in
charge of the patient. This will provide verification of POC values or initiate
exchange of the Inform monitor by nursing personnel to ensure quality patient
care.
A number of different studies have
examined the accuracy of bedside glucose meters in various patient populations
with either hypotension, poor peripheral perfusion, or edema. Two studies done
on patients in the emergency department and ICU with hypotension (systolic
blood pressure less than 80 mm Hg) both found that capillary glucose
measurement systematically underestimated venous glucose in this patient
population. (1,2) In contrast, another study performed on 75 patients with
systemic hypoperfusion (defined as systolic blood pressure less than 90 mm Hg
or requirement for vasoactive agents) found relatively good agreement between
capillary whole blood and arterial whole blood glucose, with 95% limits of agreement
of approximately [+ or -]30 mg/dL. (3) Finally, a recent study comparing
capillary whole blood to venous plasma glucose in patients with poor peripheral
perfusion related to vasopressor use or peripheral edema found that in both
categories of patients (poor perfusion and peripheral edema), capillary whole
blood glucose systematically overestimated venous plasma glucose. Even in this
study, however, almost all results agreed within approximately 2 mmol/L (36
mg/dL) glucose. (4)
In recent years, numerous studies have
been published analyzing the accuracy of glucometers specifically in the
setting of neonatal hypoglycemia. Ho, et
al. (5) reported on the sensitivity and negative predictive value of 5
glucometers in detecting neonatal hypoglycemia. They found that not even 1 of
the 5 was able to meet the ADA standards, whereas 2 of the devices were able to
meet the NCCLS standards. Khan, et al.
(6) compared 7 glucometers and reported agreement between glucometer readings
in the hypoglycemic range but found wide discrepancy in the correlation between
reference and POCT devices both in the hypo and hyperglycemic range to the tune
of 60%. The study by Ngerncham, et al.
(7) appearing in this volume of the journal too has used an elegant split
sample design and meticulously compared OneTouch SureStep Hospital Test Strips
(photometric glucose oxidase system) with a Nova StatStrip (modified glucose
oxidase based amperometric system) using Roche Modular P 800 for the reference
laboratory measurement. Another recent study reported good correlation as well
as recommended their use in neonatal clinical practice (8). All these studies
highlight that POC devices may be used as screening devices for neonatal
hypoglycemia, but confirmation of hypoglycemia with laboratory measurement of
plasma glucose is still crucial.
It needs to be realized that
laboratory estimation too is fraught with preanalytical (sample collection,
transport and physiological factors) errors. POCT devices are prone for both
pre-analytical and analytical errors (precision of the device being used). The
comparison of POCT devices with laboratory sample which has been poorly
processed can lead to erroneous estimation of discrepancy where little or none
may exist. Any further studies on comparison of POCT devices with reference
standard should focus on comparing improved second generation POCT devices with
a reference laboratory device taking utmost care to ensure precise laboratory
estimation without any processing delays and fall in glucose secondary to
glycolysis.
Australian Diabetes Educators
Association
The Australian Diabetes Educators
Association (ADEA) promotes the reliable and accurate use of blood glucose
meters within the health care setting and in diabetes self management.
ADEA recommends:
Blood glucose meters not be used as
a method of screening for diabetes.
Testing at the point of care may be
appropriate in defined circumstances, such as in remote indigenous communities
where laboratory testing is unavailable and postponement of treatment would be
potentially harmful for the individual. The meter used at the point of care
must have a vigorous quality improvement program in use.
Outside of the acute clinical
setting, blood glucose meters only be used to monitor blood glucose levels in
people with a confirmed diagnosis of diabetes.
Appropriateness of self blood
glucose monitoring be assessed on an individual basis, taking into
consideration the person’s disease and co-morbidity status, age, culture,
dexterity and physical and intellectual capabilities, identified glycaemic
targets, current medication regimen, potential confounders that may interfere
with the accuracy of results obtained, and level of motivation.
All people with diabetes using
insulin therapy are encouraged to perform blood glucose monitoring.
Individuals using blood glucose
meters have access to a health professional deemed competent in the use of the
meter.
Further diabetes education is
essential to ensure self management behaviours are underpinned by a sound
knowledge of how dietary intake, physical activity, medication, stress and
illness all interact to affect blood glucose levels.
Although some blood glucose meters
allow the measurement of glucose levels from small samples of blood from the
forearm and other sites, blood taken from the capillary bed of the fingertip is
the preferred sample, particularly when blood glucose levels are changing
rapidly.
Health professionals only use blood
glucose meters after successfully completing an education program that results
in the attainment of competency in meter operation, control testing and problem
solving.
Quality improvement practices are
implemented within clinical settings to ensure both blood glucose monitoring
equipment and operators meet high standards of performance and process.
All health services using blood
glucose meters provide:
Ø
a
well-defined policy and procedure
Ø
a
training program for personnel performing the tests
Ø
quality
improvement procedures
Ø
regular
equipment maintenance
Ø
external
auditing of meters
Ø
appropriate
lancet devices that meet infection control considerations.
In certain clinical situations
different strip technology may be more appropriate than others, such as in the
care of patients using icodextrin dialysate solution, or patients receiving
intravenous preparations containing maltose. In these situations it is
essential to consult product information and/or the manufacturer of the glucose
meter and test strip to ensure icodextrin and maltose do not interfere with
blood glucose readings.
Blood glucose measurement using
portable blood glucose meters not be used in isolation when evaluating the
glycaemic control of people with diabetes.
Substances or Conditions That May
Contribute to Total Analytical Error in Consumer and Hospital Glucose Meters
Interferant Reference
ü
Hematocrit
ü
Ascorbate
ü
Maltose
ü
Galactose
or Xylose
ü
Acetaminophen
ü
Partial
Pressure of Oxygen
ü
Dopamine
ü
Mannitol
ü
Altitude
or hypoxia
Numerous studies have been published
reporting the performance of blood glucose monitoring systems. The results of
these studies are often conflicting. Correct evaluation is, however, complex,
and the apparent contradiction of results creates confusion. One recent article
(9) provides an overview of frequently made errors in the evaluation of devices
and develops an easy-to-use checklist to verify the quality of such studies.
This checklist was used with 20 representative studies selected from the
literature between 2007 and 2012. The process revealed that limitations on the
designs and methods of studies assessing the performance of blood glucose
systems are common.
The use of the accuracy checklist showed that 40% of the studies showed clear nonconcordance with ISO 15197. The use of the interference checklist showed that only half of publications were in good agreement with the quality checks. The author concluded that many evaluations are performed poorly, and often present questionable conclusions due to the fact that few publications adhere to international guidelines. Others have also assessed the quality of glucose monitor studies and came to similar conclusions.
In the article by Mahoney and Ellison,(10)52 reports were reviewed, none of which conformed to all 38 STARD and CLSI recommendations. Future studies evaluating glucose monitoring systems should be carefully designed and should follow published recommendations to assess the quality of glucose measuring devices
The use of the accuracy checklist showed that 40% of the studies showed clear nonconcordance with ISO 15197. The use of the interference checklist showed that only half of publications were in good agreement with the quality checks. The author concluded that many evaluations are performed poorly, and often present questionable conclusions due to the fact that few publications adhere to international guidelines. Others have also assessed the quality of glucose monitor studies and came to similar conclusions.
In the article by Mahoney and Ellison,(10)52 reports were reviewed, none of which conformed to all 38 STARD and CLSI recommendations. Future studies evaluating glucose monitoring systems should be carefully designed and should follow published recommendations to assess the quality of glucose measuring devices
References:
1. Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S. Fingerstick glucose determination in shock. Ann Intern Med. 1991;114:1020-1024.
2. Sylvain HF, et al. Accuracy of fingerstick glucose values in shock patients. Am J Crit Care Med 1995;4:44-48.
3. Kulkarni A, et al. Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med. 2005;31:142-145.
4. Kanji S, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Am J Crit Care Med. 2005;33:2778-2785.
1. Atkin SH, Dasmahapatra A, Jaker MA, Chorost MI, Reddy S. Fingerstick glucose determination in shock. Ann Intern Med. 1991;114:1020-1024.
2. Sylvain HF, et al. Accuracy of fingerstick glucose values in shock patients. Am J Crit Care Med 1995;4:44-48.
3. Kulkarni A, et al. Analysis of blood glucose measurements using capillary and arterial blood samples in intensive care patients. Intensive Care Med. 2005;31:142-145.
4. Kanji S, et al. Reliability of point-of-care testing for glucose measurement in critically ill adults. Am J Crit Care Med. 2005;33:2778-2785.
5. Ho HT, Yeung WK, Young BW.
Evaluation of point of care devices in the measurement of low blood glucose in
neonatal practice. Arch Dis Child Fetal Neonatal Ed. 2004;89:F356-9.
6. Khan AI, Vasquez, Y, Gray J, Wians
FH, Kroll MH. The variability of results between point of care testing glucose
meters and the central laboratory analyzer. Arch Pathol Lab Med. 2006; 130:
1527-32.
7. Ngerncham S, Piriyanimit S,
Wongsiridej P, Inchgarm L, Kitsommart R, Vutrapongwatana P, et al. Validity of two point of care
glucometers in the diagnosis of neonatal hypoglycemia. Indian Pediatr. 2012;49:
621-5.
8. Makaya T, Memmott A, Bustani P. Point-of-care glucose monitoring on
the neonatal unit. J Pediatr Child Health. 2012;48:342-6.
9. Thorpe GH. Assessing the quality of
publications evaluating the accuracy of blood glucose monitoring systems. Diabetes
Technol Ther. 2013;15(3):253-259.
10. Mahoney J, Ellison J. Assessing
the quality of glucose monitor studies: a critical evaluation of published
reports. Clin Chem. 2007;53(6):1122-1128.
This Offer Only For US people
ReplyDeleteOne more chance to Win a Diabetes Meter- US from win-a-diabetes-meter-us.tk !
Hello Guys some days ago I win amazing diabetes-meter.It is easy to win. Just inter your email and then follow one more steps to win.It is 100% secure and relable. what are you waiting for?
SO Click Here To Win
As an increasingly important part of the healthcare system, over-the-counter (OTC) drugs are defined by the FDA as safe and effective drugs for public use without the need to seek treatment from health professionals. OTC drugs generally have the following characteristics: OTC testing
ReplyDeletearticle was informative . I agree with a lot of the points you made in this Glucose Meter article.
ReplyDelete