QUALITY CONTROL PLAN IN
PoCT
Worldwide,
one of the fastest growing aspects of clinical laboratory testing is point of
care testing (POCT), estimated to be increasing at least 10-12% per year
overall and upwards to 30% per year in some testing areas. In contrast, central
laboratory testing has grown approximately 6-7% annually.) When first widely
introduced, POCT was largely for home use or physician office laboratory (POL)
testing. In hospitals, it was considered as supplementary to central laboratory
testing. It was not generally regarded as a primary responsibility of
centralized pathology services and was often treated by central hospital
laboratories with indifference, benign neglect, or frank hostility, and
considered as substandard or second tier testing that was unmanageable. Some
laboratorians considered POCT to be a potentially disruptive competitor to
their services. POCT was also an added responsibility that many laboratories or
nursing services found difficult to assume. The attitude of POCT as an inferior
stepchild or orphan testing has changed with government regulations, the growth
of the technology, an expanding perspective and spectrum of healthcare
services, and different expectations from healthcare providers and consumers.
Decentralized patient care and access to testing in under-served areas are key
elements in the evolving expansion of POCT. POCT continues to mature both as a
technology and in the eyes of healthcare providers, laboratorians, regulators,
administrators, and vendors. While the technology has become more varied and
robust and performance has improved, the various groups associated with POCT
have grown more realistic and demanding about its potential. No longer does the
Everest theory prevail — just because it exists does not mean POCT should be
used in all situations. The need for comparability with central laboratory
testing, efficacy, operational device and kit fail-safes, management and
oversight requirements, operator performance standards, economic indicators,
and patient outcome data are all now considerations when deciding whether to
employ POCT in specific situations.
Point-of-care
testing (POCT), or bed-side testing is defined as medical testing at or near
the site of patient care. These are simple medical blood tests which can be
performed at the bedside. Simple tests such as those found in medical
examinations, urine test strips and even simple imaging such as with a portable
ultrasound device. As well as regular observations such as ECG's, O2
saturation and heart rate.
The driving
notion behind POCT is to bring the test conveniently and immediately to the
patient. This increases the likelihood that the patient, physician, and care
team will receive the results quicker, which allows for immediate clinical
management decisions to be made. POCT includes: blood glucose testing, blood
gas and electrolytes analysis, rapid coagulation testing (PT/INR,Alere,
Microvisk Ltd), rapid cardiac markers diagnostic (TRIAGE,Alere), drugs of abuse
screening, urine strips testing, pregnancy testing, fecal occult blood
analysis, food pathogens screening, hemoglobin diagnostics (HemoCue),
infectious disease testing and cholesterol screening.
The
Only Reason To Perform POCT
v Make
diagnostic decisions
v Make
treatment decisions
v Improve the
efficiency of clinical operations
Benefits
of POCT
An undisputable merit ascribed to POCT
is the ability to produce and release analytical results earlier and easier
than before by reducing the turn around time (TAT). This clear benefit, though,
has to be balanced against a number of several potential risks. Moreover, “the
faster the better” does not represent one and for all the best choice in any
setting: timeliness depends on the specific clinical situation to be managed.
The final decision should be made striking a balance among all the possible
aspects, case by case. Often, a careful analysis of the whole process will lead
to maintain a centralized diagnostics rather than to implement POCT. As a rule
of thumb, it should be accepted that POCT is redundant whenever the laboratory
can release a result as timely as clinically required. Taking into
consideration that POCT is usually much more expensive than a centralized
structure, its abuse always implies cost increases, not to mention the staff,
whose training and management is also expensive: strangely enough this item is
often ignored when preparing in-hospital POCT budgets. So it is mandatory to
demonstrate that cost increases are fully counterbalanced by a much more
efficient use of limited healthcare resources. Today, by virtue of spectacular
progress in technology and detailed standards and regulations, POCT is much
safer than a (even recent) past. Still, potential risks remain: staff with
inadequate training, insufficient supervision, and lack of system governance
and accreditation schemes. Again, it should be firmly stated that POCT can
never be considered equivalent, rather an additional tool, for a clinical
laboratory.
The variety of point of care tests
(POCT) has evolved significantly in recent years. Whilst these instruments
bring undisputed benefits in obtaining rapid results at the patient’s bedside,
these benefits are only true if the results are accurate and reliable. Quality
Controls (QC) exists to ensure accuracy and reliability.
For many of the health care workers
using POCT instruments, QC will be unfamiliar territory. Many of the standard
QC procedures applied in laboratories cannot be applied to POCT devices. However, it is essential that both primary
and community care settings apply well-structured QC procedures to ensure the
accuracy and reliability of results, minimizing risk to patients and improving
patient outcomes.
Point-of-care testing (POCT) is a
complex system with many opportunities for error. Delivering quality POCT requires
multidisciplinary coordination and an understanding of the preanalytic,
analytic, and postanalytic processes that are necessary to deliver a test
result and take clinical action. Most errors in laboratory testing occur in the
pre and postanalytical phases and many mistakes that are referred to as lab
error are actually due to poor communication, actions by others involved in the
testing process, or poorly designed processes outside the laboratory's control.
POCT requires significant operator interaction with analysis and documentation
of calibration and quality control, unlike other medical devices. Clinicians
often interpret POCT as equivalent to core laboratory testing, only faster, and
mistakenly utilize the results interchangeably despite the differences in test
methodologies. Taking quality of POCT to the next level involves looking beyond
the analytical phase and integration of POCT into the entire pathway of patient
care to understand how POCT relates to medical decision-making at specific
points during the patient's care.
In recognition of this there is now an
international standard for POCT, namely ISO 22870 which, in conjunction with
ISO 15189, lays out the requirements for quality and competence. However, there is no ‘one-size-fits-all’
solution.
Quality Control for POCT
Electronic quality control: EQC checks the electronics of the device but not the
reagents or cartridge.
Auto-QC: Usually used on multi-use cartridge type devices especially blood gas
analyzers. QC is performed automatically and verifed by the software on the
instrument.
Built in QC: Usually used on single use visually read strips or
instrument-read cartridges. Uses a control band that turns positive if the
device is working correctly.
Developing an appropriate QC plan
for POCT
Firstly, the actual design of the POCT
device needs to be factored in. These broadly fall into 3 categories:
Laboratory type instruments – full size instruments, for example
blood gas analysers. Similar to instruments you might find in a laboratory, the
QC procedures for these types of analysers
should follow full laboratory QC protocol. Multi-level QC samples should be run
on these instruments every day a patient test is performed and the accuracy and
reliability of those results should be monitored over time by participating in
a frequent PT scheme.
Cartridge-based instruments – for example HbA1c and INR
analysers. These are usually very different from that found in a standard
laboratory, consisting of a cartridge based component and an electronic reader,
which may have a self-check system built in. The cartridge contains all the
necessary components for the analysis of the patient sample while the
electronic reader is responsible for converting the results from the cartridge
component into a readerable numerical value.
The difficulty with QC on
cartridge-based instruments is that you can only ever test that one particular
disposable cartridge and the electronic functioning of the instrument. However,
for such devices QC is still essential. The cartridge may have been damaged
during transit or the on-board reagents may have deteriorated over time. It is
therefore recommended, as a minimum, to run QC when changing cartridge lot and
periodically throughout the lifespan of the lot to ensure the stability of the
on-board components. It would also be beneficial to participate in a frequent
PT scheme to monitor the accuracy of reporting over time.
Strip based instruments – for example electrochemical or
reflectance strip based glucose meters. These are similar to cartridge based
instruments in that the strips are responsible for the analysis of the sample.
However, unlike cartridge devices the electronic component has no self-check
feature and without this a faulty analyser could be producing erroneous results
for some time undetected. This makes QC even more important for these types of
instruments. Strips should be checked using multi-level QC on delivery and
every day of patient testing. Liquid-ready
multi-analyte third party quality controls available from some QC
manufacturers are ideally suited for this as they require no advance
preparation and are easy for non-laboratory staff to use.
Regular proficiency testing is also
important for these devices to ensure accuracy of reporting over time.
Proper Quality Control Reagents
Accuracy assessments are performed to
verify that a system functions properly. Some manufacturers would recommend the
use of liquid controls as a quality check. However, the composition of these
liquid controls (also known as quality control reagents) should be as similar
to whole blood as possible. Solutions such as red dye are not acceptable
according to established ICSH standards. There is something known as the
“Matrix effect” which is a term used to describe the differing performance
characteristics between a liquid control and patient blood. This difference is
hardly identifiable and almost impossible to trace reliably due to the complex
chemical, physical, and biochemical interactions. Thus, even a Proficiency
Testing (PT) and an external quality assessment (EQA) can yield inaccurate
conclusions unless it uses whole fresh blood. The ICSH also recommends devices
that use whole blood count instead of diluted blood.
Traceability
Traceability refers to the chain of
referencing an instrument has. The ability of a manufacturer to say that their
reference instruments for their systems have been calibrated against the
international reference method established by the ICSH or by the National
Institute of Standards and Technology is significant. It is plainly significant
because just like other issues of standardisation, the user needs to know what
to expect based on what is sold to them. In the medical field, standardisation
has to do with life and the sustaining of life. There can be no compromise of
extreme accuracy in this field. Therefore, having a standard reference that is
traceable would also mean someone has to be accountable; and should there be
issues raised or contrary findings appear, an already established platform is
there to address it as a global community.
How
to implement an in-hospital POCT
system
First and foremost, the set up of a
MSC, appointed by the top Management of the Institution to whom devolve power
for:
²
sharing
the project with all stakeholders;
²
representing
in full the Authority presiding over the in-hospital
POCT system;
²
defining
competences and responsibilities for each component inside the project;
²
defining
criteria for preliminary choices before starting with POCT;
²
carefully
appraising the available resources;
²
selecting
wards/services where POCT devices are to be placed;
²
selecting
a test menu;
²
activating
a regular scheme of clinical audit about the appropriateness of the whole
organization in order to check:
²
system
effectiveness, based on measurable outcomes;
²
system
soundness;
²
evaluating
on a regular basis the global cost/effectiveness;
²
planning
continuous quality improvement of the system.
Who is responsible for POCT –
clinicians or laboratory staff?
Responsibility for QC on POCT
instruments, although located in the clinical setting, should ultimately lie
with the laboratory. However it should be a team effort involving the staff who
are using the actual equipment. Training of staff to use the POCT equipment,
including QC, and the on-going competency of operators should be documented and
governed as part of an overall quality management programme. Training staff who
are using the POCT equipment to perform QC will give them a better
understanding of how the equipment works and in doing so will help them to see
for themselves when errors occur. However, ultimately the responsibility of
quality control and should lie with the laboratory.
Disadvantages Of POCT
•Higher unit cost: Conventional wisdom
POCT is more expensive than central laboratory testing
•Difficult to manage within in
regulatory requirements
•Higher rate of operator errors in
specimen collection, test performance, data management and documentation for
regulatory compliance
GUIDELINES FOR CONDUCTING QUALITY CONTROL AND EXTERNAL QUALITY
ASSURANCE FOR PoCT (AUSTRALASIAN ASSOCIATION OF CLINICAL BIOCHEMISTS )
1. QUALITY CONTROL
(QC)
Minimum requirements
for running quality control: QC shall be performed on all PoCT devices.
There are two main types of PoCT devices: low and medium
complexity. The latter category includes blood gas devices that have a single
use cartridge. Most PoCT devices have a large number of in-built quality
checks.
For low complexity devices
utilising strip technology (eg glucose or coagulation meters), a minimum of one
liquid quality control (QC) sample shall be tested each month unless a higher
frequency is suggested by the manufacturer. It is recommended that if only one
QC sample is run it should have a concentration in the clinically relevant
range for the analyte being measured. If two levels of QC are available, then
control samples with both a normal and an abnormal level should be run.
Medium complexity devices utilising
cartridge based technology (eg i-STAT, DCA) shall have a minimum of two liquid
QC samples tested each month unless a higher frequency is suggested by the
manufacturer. These two QC samples should contain normal and abnormal levels.
In addition to the
regular QC program, QC testing should also be undertaken when:
v The lot number of
consumables changes
v There is a new
delivery of consumables
v An operator lacks
confidence in a patient result
v The health care
professional does not believe that the PoCT result fits the patient’s clinical
picture
v Substantial
maintenance procedures have been carried out on the device.
v The device has suffered a
physical insult (eg dropped, temperature extremes – hot or cold, etc)
Electronic QC is a check of the device’s measurement signal only
and does not check the analytical part of the system. Therefore it is
complimentary to liquid QC requirements and not a substitute for the minimum QC
requirements outlined in this document.
Who should run PoCT
quality control samples?
QC testing for PoCT shall be undertaken by the PoCT Operator. All
operators who at any time use the device should participate in the quality
control program.
Who should be
responsible for training PoCT Operators to test quality control?
Ideally there shall be a collaborative approach between PoCT
Coordinators, suppliers and if appropriate the associated NATA/RCPA accredited
laboratory. Minimum acceptable operator training standards shall be documented
and available for review.
At completion of training, PoCT Operators should have their
knowledge and technical competency assessed.
Who should be
responsible for reviewing quality control results?
The PoCT Operator undertaking QC testing shall immediately review
the QC results and take any appropriate action.
The PoCT Coordinator or delegate should be responsible for review
and trend analysis of QC results, taking appropriate corrective action when
required and operator participation. All QC records, reviews and corrective
action documentation should be maintained for three years and be available for
review.
2. EXTERNAL QUALITY
ASSURANCE (EQA)
A form of external quality assurance should be undertaken for
every PoCT device.
Participation in a recognised EQA program is recommended for each
analyte being tested. Monthly split patient sample testing with an accredited
laboratory may be considered as an alternative form of external quality
assurance. This mode of EQA may be useful when a commercial program is either
unavailable or unsuitable for the instrument in question.
Note: Users must be
aware of limitations of using the split patient sample approach which include
laboratory method performance, limited range of testing, transport stability
and lack of peer comparison.
Who should run PoCT
EQA samples?
EQA for PoCT should be run by the PoCT Operator.
Who should be
responsible for training PoCT Operators to test EQA samples?
Ideally there should be a collaborative approach between PoCT
Coordinators, EQA Scheme Organisers, suppliers and if appropriate the
associated NATA/RCPA accredited laboratory. Minimal acceptable training
standards, including refresher training should be documented and available for
review.
At completion of training, PoCT Operators shall have their knowledge
and technical competency assessed.
Who should be
responsible for reviewing EQA results?
The PoCT Coordinator
or delegate should be responsible for the review of EQA results and taking
appropriate action. All EQA records, reviews and documentation should be
available for review and maintained for three years.
POCT Quality Control Guidelines
TEST
SUPPLIES AND CONTROL MATERIALS:
²
No
expired Controls or Testing supplies.
²
Proper
storage of controls and test materials. When Awarepoint temperature monitoring system
is not available, Temperature Logs monitored in pyxis, rooms and refrigerators
are required.
²
Testing
reagents and supplies QC’d, dated (month/day/year) and initialed.
²
No
loose supplies (Colo./Gast. Slides; Pregnancy Test kits). Test slides must be
kept in the original box.
²
Control
materials dated (month/day/year) and initialed. Write discard date
(month/day/year) if shelf-life is based on open date; eg. glucose control
solution expires in 90 days upon initial opening or the expiration date printed
on the vial if it comes first.
QC
LOGS:
²
QC
frequency performed according to POCT policies and procedure.
²
Document
all QC results on the QC logs.
²
Verify
numeric QC results are within the numeric range noted on the QC log; e.g.
Multistix QC pH result within QC range on log sheet. Follow QC log result
manner of reporting.
²
Write
“POS” for Positive and “NEG” for negative. Do not use ( + ) or ( - ) signs.
²
Document
“Not In Use” (NIU) on days when patient testing not performed and QC not
required.
²
Document
Corrective Actions for FAILED QCs.
²
Control
Materials and Test Supply lot #s must match lot # on QC log.
²
Use
black or blue ink pens for QC documentations. Do not use pencil or red ink
pens.
²
Error
Corrections:
v
Draw
a line through error, Initials and Write correct results.
v
No
“Whiteout”
v
No
“Write Over”
v
No
“Xerox copies”
v
Document
all equipment maintenance.
v
Managers/Supervisors
must review and sign “Monthly Manager/Supervisor
Reviews” section of QC log.
v
QC
logs are kept for at least 3 years as required by law.
COMPETENCY:
Records of staff annual competency
must be available.
No comments:
Post a Comment