AUTOVERIFICATION IN CLINICAL LABORATORY
The
automatic release of results from clinical instruments via algorithms running
in a laboratory information system (LIS) may improve efficiency, reduce overall
turnaround time, and be accomplished within current regulatory frameworks. Autoverification is a process
whereby clinical laboratory results are released without manual human
intervention. Autoverification uses predefined computer rules to govern release
of results. Autoverification rules may include decisions based on instrument
error flags (e.g. short sample, possible bubbles, or clot), interference
indices (e.g. hemolysis, icterus, and lipemia), reference ranges, analytical
measurement range (AMR), critical values, and delta checks (comparison of
current value to previous values, if available,from the same patient). Rules
may also define potentially absurd (physiologically improbable) values for some
analytes and additionally may control automated dilutions and conditions for
repeat analysis of specimens. More sophisticated application of autoverification
rules can generate customized interpretive text based on patterns of laboratory
values. Autoverification is commonly performed using the laboratory information
system (LIS) and/or middleware software that resides between the laboratory
instruments and the LIS. Autoverification can greatly reduce manual review time
and effort by laboratory staff, limiting staff screen fatigue caused by
reviewing and verifying hundreds to thousands of results per shift. Ideally,
autoverification allows laboratorystaff to focus manual review on a small
portion of potentially problematic specimens and test results. However,
improperly designed autoverification can lead to release of results that should
have been held, potentially negatively impacting patient management. There is a
guideline document produced by the Clinical and Laboratory Standards Institute
(CLSI) on autoverification of clinical laboratory test results which focuses on
the process for validating and implementing autoverification protocols-CLSI AUTO 10-A; Volume 26 Number 4.
The
autoverification rules evolved over more than a decade, with a steady increase
in autoverification rate to the current rate of 99.5%. The high rate of
autoverification is driven in large part by the highest volume tests or test panels
(e.g. basic metabolic panel, albumin, alanine aminotransferase, and troponin
T), which all have autoverification rates exceeding 99.0%. This frees up staff
time to deal with assays such as certain drug levelsor endocrinology tests that
require offline steps such as manual dilutions, or to investigate questionable
test results. Some tests in our study currently have autoverification rates
under 90%; however, these tests comprise a small fraction of the total test
volume. Informatics support is critical to successful implementation and
maintenance of autoverification. The most common problems interfering with
autoverification would be interruptions of network, LIS, middleware, and/or the
interfaces between these systems.
The
biggest risk associated with this type of process is releasing large numbers of
results without proper review or editing. This results from poor planning,
implementation, or failure to follow procedures. A lesser risk is to not
release results that do meet the autoverification procedure. Other than affecting
turnaround time and workflow, the latter is a “safe” failure. Any autofiling
procedures or modifications to software should not interfere with or inactivate
the normal procedures and functions of either the instrument or LIS. To as
large an extent as possible, the process should be fault tolerant of user mistakes.
The worst that should happen is that results are not released when they could
have been released. The process should be tested thoroughly with all
possibilities of instrument flags, errors, and ranges and it should also
periodically be revalidated. During testing and validation, each criterion for autofiling
needs to have an individual example tested and determined that it performs as
expected. That is, the system releases data when it is supposed to release and
held when it is supposed to be held. Combinations of criteria need to be
considered in addition to the order that results are released by an analyzer.
Ultimately, the process will only be as good as the planning and testing.
The
Joint Commission on Accreditation of Healthcare Organizations (JCAHO) does not
appear to have standards that are directly applicable to autoverification.
Standards under quality control that may apply indirectly include QC 1.3 and QC
1.4. QC 1.3 states, “The laboratory’s quality control system includes daily
surveillance of results by appropriate personnel.” The description of the
intent of this standard include statements that “a computer may be used to
screen results, using similar specific criteria, so that only outliers need to
be reviewed manually.” QC 1.4 states, “The laboratory takes remedial action for
deficiencies identified through quality control measures or authorized
inspections and documents such actions.” Autoverification must meet the intent
of this standard by incorporating quality control into the process. A LIS
requirement for autofiling must include the ability to use bar coded sample
specimens. These are ubiquitous in laboratories today although there may still
be a small subset of samples that are labeled without bar codes. The bar coded
or otherwise machine readable label allows the positive identification of the
sample. This is necessary so that correct criteria, particularly delta checking
information, are applied by the autoverification algorithms. It is important to
have positive identification of a specimen. An LIS that requires creating lists
of accessions, especially manually created lists, and then applying an
algorithm to the list in sequence introduces too many chances for mix-up or
phase shifts.
The
LIS software must have some facility or function for autofiling as a feature.
For regulatory reasons, laboratory personnel must work within the constraints
of the software or vendor. Modifications to instruments or software can
introduce the possibility of FDA oversight. Some vendors will perform custom
programming for a fee. This tends to be expensive and makes support and
upgrades very difficult. Finally, algorithms and programs should be designed
from the assumption of error. Criteria for release should be explicitly spelled
out. The opposite approach would be to design the release algorithm on the
assumption that all results presented by the instrument are acceptable for
release unless criteria are met that invalidates them. This is a subtle but
important distinction. For
example
suppose the data module of an analyzer flags results with a 0 if there is no
error and a 1, 2, or 3 to indicate various alert conditions. A criterion can be
defined to release results providing the instrument flag is not a 2 or 3. The
assumption is that a 0 or 1 is acceptable. But what if the instrument sends a
4, or more likely, a text character that indicates a severe instrument
malfunction? The results would be released if the criterion for release is
“release if flag is not equal to 2 or 3.” If the criterion for release is
instead “release if flag is equal to 0 or 1” then any unexpected characters in the
data stream would not result in accidental release of results. The LIS should
provide the functionality to hold or fail an entire cup (entire specimen such
as a CBC) as well as to fail individual tests (HGB only) allowing maximum
flexibility in defining criteria. The LIS needs to be capable of holding failed
cups or tests in a recheck or re-filing queue. Not every sample will be
autofiled and manual data release should take place in a normal fashion. If
samples are re-run the autoverification program should be able to display both
the original results and re-check results. Autoverification should not be
applied to re-check data. Once implemented, the process remains dynamic.
Initially the rate of autoverification should be calculated to determine if it meets
the goals established during implementation. This rate may change over time as
the patient population changes. Changes in reimbursement may affect
test-ordering patterns. As the complexity of tests ordered changes,
autoverification rates will vary. Large numbers of relatively normal patients
being screened are what make this process the most efficient. With managed
care, screening testing is becoming less frequent, patient populations may consist
of a larger percentage of “abnormal” samples, lowering the autoverification
rates. New instrument models with expanded capabilities and parameters may make
this process easier. The data handling abilities are becoming more
sophisticated with much more versatility in turning flags on and off, setting
ranges, and storing information from the LIS. Laboratory information systems are
also becoming more sophisticated with better integration of autofiling
functionality. Setting the criteria and rules will become easier with resulting
test efficiencies. Autoverification can positively affect workflow in the
laboratory.
Thanks for your post which is truly informative for us and we will surely keep visiting this website.
ReplyDeletemedical labs in kukatpally
book best doctors in vijayawada
Book Doctor appointments online at ameerpet
order blood test online at gachibowli
Total Diagnostic Centres in Hyderabad
blood test online report
Very impressive article! The blog is highly informative and has answered all my questions.
ReplyDeleteTo introduce about our company and the activities, B2B contact list is a database provider
that helps you to boost your sales & grow your business through well-build
clinical lab scientists mailing-lists .
Great information for those People who are looking for Diagnostic Centres in Madinaguda .
ReplyDeleteWe Avigna Diagnostic is proven one of the Best diagnostic services in Telangana with high and advanced laboratory with well trained technician. We provide the test results with an accurate.