Challenges in
POCT
Point of care testing (POCT)
refers to testing that is performed near or at the site of a patient with the
result leading to a possible change in the care of the patient. Although there
are many benefits of using POCT devices in terms of their convenience,
establishing a POCT indeed is a challenging job. Some of the biggest challenges
for the person holding this post are gaining physician and nurse allies and
turning non-laboratorians into testing personnel, all while ensuring adherence
to best laboratory practices and regulatory agency standards. In many hospitals
a range of tests are performed in the wards, by staff who are totally
uneducated in laboratory sciences and who may not even have the relevant
skillsin quality assessment to carry out the tests competently. In this article
we have review the challenges that needs to be overcome while setting up a
POCT.
POCT: Challenges
l Management
l Responsibility
l Location
l Staff training and competency
maintenance
l Reliability of POCT results
l Quality Control
l POCT Challenges for End Users:
l Data Management
l Noncompliance with procedures (specimen
labeling, QC, proficiency testing etc.)
l Infection Control
l Billing
II. MANAGEMENT OF POINT-OF-CARE IN THE HOSPITAL
Management of POCT is
challenging –there can be dozens of sites, hundreds of POCT devices/kits, and
thousands of operators to manage to assure quality of testing.
The first challenge
in developing a strategy to manage POCT involves building a competent
interdisciplinary POC management team including the laboratory, physicians, and
nurses. The POC team should hold the ultimate responsibility for determining
the test menu, selecting technologies, establishing policies and procedures,
ensuring training and regulatory compliance, and providing advisory assistance
to the end users of POC technologies. After establishing a POC team a
management structure should be build that is responsible to implement new
initiatives and to perform corrective action where necessary.
The role of the POCT
team includes the following(1):
Ø determining if POCT
is justified at a particular location. This would include a clear demonstration
of increased clinical effectiveness
Ø establishing a system
for the continuing audit and assessmentof POCT
Ø ensuring that no POCT
device is used unless it has been looked at by the POCT committee
Ø setting up a quality
hierarchy to ensure that there is a direct link between the person performing
the analysis and the POCT committee
Ø establishing
the presence of a link nurse or other healthcare professional at the point of
service delivery
The POC management
team serves in a consulting role and coordinates the POC program throughout the
institution.
Ø including
representatives from primary care and the community where necessary
Ø ensuring that users
have documented training in the use of POCT devices and that they are fully
aware of all contra-indications and limitations
Ø ensuringthat
internal quality control (IQC) and external quality assessment (EQA) schemes
are applied to POCT in the same way as they would be for the established
laboratory service.
Managers of POCT
should also be aware of their responsibility for clinical governance and of the
medico-legal implications of an erroneous result.
Lines of
accountability should be clearly written into local policies and procedures and
should cover the following areas:
Ø training
Ø instructions for use
Ø standard operating
procedures
Ø health and safety
Ø quality assurance
Ø maintenance
Ø accreditation
Ø record keeping
Ø audit
Ø adverse
incident reporting.
III. WHOSE RESPONSIBILITY?
Who holds the
responsible for the test results performed outside the laboratory -the
operator, the laboratory or the manufacturer? The rational answer to this
questionwould be: all three parties must accept their responsibility toward
assuring
the accuracy of every single result. Clinician is the person who is going to
take the decision on patients condition whether the test is performed in the
laboratory or bedside, whatever the
results are obtained expected or unexpected regarding the treatment or
repeating test or further work up that may be necessary to arrive at the right
diagnosis. The clinician needs to establish his confidence on the reliability
of the results obtained out of POCT(2). In reality, the manufacturer cannot
monitor a hospital to ensure that all the recommended procedures are being followed
and that documentation is adequate. The place where manufacture can contribute
is by provideing quality instrument and by performing planned-unplanned
maintenance and calibration on routine basis. All these measures do not implies
that the laboratory has been relieved of all responsibilities for the quality
of these POCT results.(3) The central laboratory, when appropriate, should be
responsible for evaluating each new piece of instrument and each newlot of
reagent, and maintaining a performance record, carrying out co-relation studies
between laboratory instrument and POC instrument. Tasks and responsibilities
can be moved across traditional territories. By identifying where the process
crosses territories, opportunities for cooperating and adopting a total system
perspective can lead to powerful new solutions to common problems. Typically,
the cross-territory concept occurs in three areas of the hospital, i.e. the
clinical user unit, the laboratory and the information technology department
(ITD).(4) In order to improve the utilization of POCT and the quality of
patient care, and to ensure that results are integrated into and being
networked with the laboratory information system, the establishment of new
relationships among the laboratory, clinicians and the ITD people is needed.(5)
IV. TRAINING USERS IN POCT
Should cover theory
and practice:
–Storage of QC
material
–Preparation of QC
material
–Appropriate QC
frequency for each test
–Principles and
practice of QC testing
–Remedial actions for
failures
–Recording and
documentation of actions
l Once the procedure
manual has been created it is necessary that all users have read it and this
process should be documented.
l The salient features
for each test need to be understood and demonstrated during training.
l This
would be followed by a competency assessment which would be the exit
examination before the healthcare worker can test patients.
V. CHALLENGES IN TRAINING
Management
of Training
v Number ofstaff
requiring training be in the thousands
v A high turnover rate
v Nursing staff can be
transferred between different departments
v Poor communication
between nursing administration and POC coordinator.
v Diverse
educational backgrounds
The
Designated Trainer
v Better way to train
new hires
v This
designated trainer would aware of all the issues that could occur with the test
and would be up-to-date in their competencies
Train
the Trainer (6)
v Determine who will
conduct the training
v Ensure that this
person(s) has the appropriate qualifications, experience and training to be
able to train others and that these are documented in their training records
v Conduct the trainer
training, and assess their competence to provide training to others
v Document
the training and competency assessment.
Peer
To Peer Training
v Nurse trainer may not
be aware of all the preanalytical, analytical and post-analytical errors that
could occur when using the POC test.
v They may also be
performing the test in a different way that could be prone toa higher error
rate.
v Furthermore,
from a regulatory standpoint, the trainer would need to have passed their
competencies as well.
Training
Methods
v Direct classroom
demonstrations and observations
v Supplemented
with self-learning , e-learning, lectures via webcam, power point, and a
training kit containing manual, laminated posters/aids and CDROM.
Competency
Assessment
The aim: To ensure
healthcare workers can not only generate quality results consistently from the
instrument but can correctly manage them inthe decision making process.
Ø Practical Assessment
Ø Written
Assessment
VI. NON COMPLIANCE WITH PROCEDURES
Majority of the staff
involved in the POCT is from non-laboratory side which is unaware of the
routine laboratory procedure regarding calibration, maintenance of instruments,
quality control processing and analysis of QC data, sample collection procedure
and patient preparation before sample collection which posses a threat of
non-compliance with standard operating procedure that may contribute to source
of errors in POCT. These staff are not aware of the interfering substance,
effect of hemolysis and lipemia or compromise samples which results into
analytical error or spurious results of the test. Not following the standard
operating procedure may contribute to analytical error in addition to pre
analytical errors. These are the some of the challenges in case of
non-compliance with the procedure which can be eliminated through training and
retraining of POC staff and awareness programme.
VII. REALIABILITY OF POCT RESULTS
Questionable quality
can occur, given the variety of educational and experience levels and turnover
of staff that perform the tests. Greater inter-individual variability in
results (compared to central laboratory testing) is common. Waived category
does not guarantee reliability. Simplicity is deceptive and there are many ways
that staff can inadvertently generate a wrong result with waived or “simple”
tests. POCT results are not necessarily comparable to central laboratory
results –Standard methods may not be used in POCT and thus it may not be
possible to compare results across sites (e.g. when patients travel and are
tested at different sites, or when treatment protocols derived from more
accurate results are being followed). Differences in specimen types, (e.g.
serum, plasma, or whole blood,) may also affect results between traditional
central laboratory methods and POCT. Thus, clinical protocols based on central
laboratory results may need to be revised when utilizing POCT results. POCT
kits and devices may not be FDA approved for all uses that a similar test in
the central laboratory can be used for.
VIII. QUALITY CONTROL
Running quality
controls (QC) is a mandatory requirement for any point-of-care test because
they are designed to detect problems in the test system. They are tested before
or alongside a patient test and should always be run according to the
manufacturer’s instructions in the product insert. Their use monitors test kit
and reagent integrity. For example a QC failure could arise from incorrect
storage of kits and/or also be due to poor techniques or procedures. Thus QC
gives assurance that the device is working and the testing is being performed
correctly.
IX. CHALLENGES IN POCT IQC
• Especially, when no
enhanced QC design features on the POCT device
• User dependent +++
–Users neglect to
perform QC
–Users fail to take
corrective action for out of range results.
–Users fail to
document or record QC results
-POCT operator most
of the type is non-laboratory staff which is not trained in quality control
testing and trouble shooting in case of outlier.
• Expensive costs,
especially in low volume settings (QC/patient ratio).
-QC material not
properly stored -stability problem.
Laboratory
professionals are best able to set quality testing recommendations practicable
in the field. They should tailor programs without excessive complexity adapted
to be performed by non-laboratory operators in a non-laboratory environment.
The degree of technological improvement of the PoCT devices, the level of
connectivity and the volume of patient testing are major elements to be taken
into account.
Non-instrumented
qualitative point-of-care tests such as pregnancy tests, HIV tests, rapid strep
A or flu tests typically have two types of controls:
Internal controls
-built into the test system and are run whenever a patient sample is tested.
Confirm that the test system is working, and for lateral flow methods,
sufficient specimen has been added to the well to allow the sample to migrate
correctly through the strip. The goal of QC testing is to ensure that the PoCT
system and the operator are performing correctly (testing reliability and
routine work quality) and that results correspond to the expected values of the
control material. The QC procedure includes control material testing, immediate
results analysis and identification of errors to undertake remedial actions. If
QC results fail, patient testing should not be performed until the problem is
resolved. Many devices designed for PoCT do not really fit with traditional
laboratory QC systems (strip-based devices or cartridge-based) and
new-generation devices also have in-built quality controls that automatically
check the device. For these instruments, laboratory professionals should have
an understanding of what type of instrument has to be checked and what parts of
the instrument will be checked to set up the QC program (frequency and nature
of control materials).(7)
X. CHALLENGES IN POCT EQA
• Limited period of
time to perform testing and to return results on each POCT site
• Low frequency of
testing
• Packaging: sealed
glass vials…
–How to safely break
open the vial
–How to reconstitute
the material
–How pipetting the
material if necessary
• In case of
unsuccessful result:
–identifying
the cause
–developing an action
plan
• Recording of
results
EQA is strongly
recommendedfor all point of care devices. Considering the above challenges
mentioned an EQA programme should be selected that offers:
v Frequent reporting to
minimise the amount of time an error can go unnoticed.
v Quality material in a
format suitable for use with POCT devices.
v Well-designed reports
that allow for quick and easy troubleshooting of erroneous results at a glance.
v Multiple
instrument registrations for each EQA sample provided, helping to save money
and monitor performance across all POC instruments.
XI. POCT CHALLENGES OF END USERS
Competencies -Waived
vs Non-waived
• Underestimation of
risk by the user
–False perception of
infallibility
–Pressures of a busy
clinical environment
• Training and
competency verification
–Large number of
users
–Diverse educational
backgrounds
• Technical aspects
–Many locations to
control (metrology, documents…)
–Adequate storage
space to store specimens to repeat the tests between the last successful QC and
a failed QC
–Clinical management
on receipt of results does not allow the system to be out of control.
XII. INFECTION CONTROL
Standard (universal)
infection control precautions
v the prevention of
occupational exposure to blood-borne viruses, the wearing of gloves and other
protective clothing, and the prevention of sharps injuries
v prevention of cross
infection with blood-borne viruses, including selection of appropriate lancing
devices
v safe handling and
disposal of healthcare waste, including sharps
v Hand washing is
generally considered the most important measure to prevent the spread of
infection. Hands should be washed before patient contact, after patient contact
and after contact with body (6)
v fluids irrespective
of whether gloves are worn or not
v safe
medical device use, including decontamination of reusable devices.
In an unpublished
case study outbreaks of hepatitis B were reported from several environments
where blood glucose monitoring was being carried out for multiple patients.
Thorough investigations identified that care workers were found to be using
lancing devices intended for self-use (by one patient only) to take blood
samples from multiple patients. This use of the wrong sort of lancing device
was implicated in the transmission of the virus.
XIII. POCT CHALLENGES OF THE DATA MANAGEMENT
POCT data is
available at the place of testing but once the patient is shifted form
intensive care units or emergency department to general wards or subsequent
follow up visit in OPD, the data of POC testing is not available if the devices
are not well connected with LIS and HIS. To make this reports of test available
at other sites than POC these devices needs to be connected with LIS and HIS.
POCT results in the EMR pose additional levels of complexity. Connecting POCT
devices to an EMR, for example, requires a computer interface and ongoing
maintenance at additional costs. Many POCT devices must pass results through a
proprietary data manager (from the specific manufacturer of the device) through
a laboratory information system, then into the EMR. Fortunately, universal
connectivity options are available that can connect to several POCT devices
from different manufacturers and simplifies the transfer of results from
multiple devices in an institution. Keep in mind that universal connection
creates additional steps in the sequence of passing data from the device to the
EMR.
A primary
disadvantage of current POCT data management is the requirement to periodically
connect or dock a device before the data can pass to the EMR. Data can stay in
the device for several hours, days or even weeks if the device has not been
docked. Reliance on manual docking delays documentation and billing of patient
results.
Although glucose
meters, coagulation and blood gas devices have data management and connectivity
features, nearly half of the POCT market is visually interpreted dipsticks or
kitsthat require manual data entry to document results in the EMR. Pregnancy,
rapid strep, urine dipsticks, HIV and occult blood are all visually interpreted
tests that have no connectivity options. Many of the test included are manual
dipstick test or card test, in such case it is a tedious job to enter the all
data manually which makes the process of data management a challenging task.
Manual result entry is an additional step for staff that takes them away from
their patient care responsibilities, so institutions cannot expect to achieve
100 percent compliance with documentation of every result.
Location
POCT is conducted at
many different locations and can be under different laboratory licenses and
medical directors. In a large health system with many testing locations, the
display of dozens of glucoses would be extremely confusing in the EMR. A better
option is to link this information to each result but suppress the display of
the site information such that any clinician or inspector can right-click the
mouse to find the analyzing site, address and medical director at the time the
test was conducted.
Billing
POC testingare
associated with high cost per test due to their rapidity of result generation
and consumable cost which is different from that of central laboratory testing.
This posses an additional pressure on billing system to assign different code
for these test. This again contribute to the source of error in billing in
hands of untrained or newly appointed staff resulting into failure to input
thedata into data management system due to mismatched test code. Also in case
serial testing of some parameter that are used for monitoring of patient
condition such as ABG, urine ketone, blood glucose testing the tests are
carried out frequently. In such cases there are chances of not raising the
bills for such frequent test. This will lead to financial losses to institution
as well as these unbilled test reports will not be entered into data management
system that will lead to incomplete records and subsequent impact on management
of patients condition.
XIV. CONCLUSION
POCT implementation
requires a systematic approach which involves all stakeholders. Most of the
healthcare organization are unfamiliar with the POCT due to which very less
efforts are made to establish a complete POCT set up. Predictably, issues
concerning the quality of test results became apparent resulting in the
establishment of regulatory requirements that have been enforced by hospital
accreditation organizations. Failure to comply with regulatory mandates may
have major consequences.
Thanks for sharing nice information about suitable & safe blood glucose test strips with us. i glad to read this post.
ReplyDelete